Hamilton Microlab VANTAGE

Get a Hamilton Microlab VANTAGE connected through GroundControl and confirm the connection from Python without moving the instrument.

This page covers connector installation, configuration, and a first test for the Hamilton Microlab VANTAGE line of liquid handlers. The Hamilton manuals cover instrument installation, maintenance, and safety. To operate the instrument, see the Hamilton VANTAGE guides.

What you need

ItemDetail
InstrumentA Microlab VANTAGE / STAR V 1.3 or 2.0, powered on
NetworkAn Ethernet connection between the computer and the Pipettor, and the Pipettor's IP address. Configure the computer's Ethernet adapter with a static IP address on the same subnet as the Pipettor. The Pipettor typically uses 192.168.1.20, but network configuration may vary.
Connectorunitelabs-microlab-vantage connector
Vendor softwareNone needed. The connector communicates with the instrument directly over Ethernet.
MachineWindows, Linux, or macOS running GroundControl
AccessA tenant and the SDK installed. See Setup

Models

The Microlab VANTAGE comes in 1.3 m and 2.0 m widths. A VANTAGE instrument can consist of up to three modules:

  • Pipettor: the liquid handler itself, with deck and pipetting tools mounted on one or multiple arms.
  • Logistic Cabinet: the storage space below the deck. It holds active components such as Entry/Exit modules, and space for storage or integration.
  • Track Gripper: a separate robotic arm at the back of the instrument.

The Microlab STAR V is a VANTAGE system consisting only of a benchtop Pipettor, without the Logistic Cabinet and the Track Gripper.

The unitelabs-microlab-vantage connector supports the Pipettor module. The Logistic Cabinet and the Track Gripper aren't supported.

Each module has its own IP address. An internal router connects the modules, so only one Ethernet cable leaves the instrument. Use the Pipettor's address in the connector configuration.

VANTAGE Pipettors can differ in their hardware configuration, including the presence and number of pipetting channels, pipetting heads, and transport tools. The unitelabs-microlab-vantage connector supports all models and reads their configuration from the instrument firmware. It exposes only the modules present on the physical device through the SiLA 2 interface. See Supported modules for details.

For detailed model and configuration specifications, consult the manufacturer's documentation.

Install and configure the connector

  1. In GroundControl, add a device, search for Vantage, and select the model: Microlab VANTAGE 1.3 or Microlab VANTAGE 2.0 (how to add a connector). Microlab STAR V instruments use the VANTAGE entries.
  2. Fill in the device-specific configuration:
    FieldValue
    hostThe IP address or hostname of the Pipettor. Typically 192.168.1.20. The connector default is 192.168.1.1
    portThe TCP port of the instrument. The default is 24652, which is typical for VANTAGE instruments
    command_timeoutSeconds to wait for the XL module initialization to finish. The default is 60. All other commands use a fixed timeout of 5 minutes, which this setting doesn't change
    has_core96Optional override. The CO-RE 96 head is detected automatically at startup. Set to true to register the Core 96 Service when detection fails. The default is false
  3. Start the connector. It opens a connection to the instrument and reads which modules are installed. GroundControl then lists the services for those modules.

To run the connector on a machine without GroundControl, see Headless install.

Supported modules

The UniteLabs VANTAGE connector supports the following modules. It detects which modules are present on the device and registers a service for each one.

ModuleConnector serviceRegistered whenSince connector version
Instrument coreDevice Service, Gateway Service, ARM ServiceAlways0.1.0
1000 µL channelsPIP ServiceAn arm has pipetting channels0.1.0
5 mL channels (XL)XL ServiceAn arm has XL channels0.3.0
Internal Plate Gripper (IPG)IPG ServiceAn arm has an internal plate gripper0.1.0
ID loader (IDL)IDL ServiceThe instrument has an ID loader0.1.0
Quad CO-RE Gripper (QCG)Core Gripper ServiceAn arm has pipetting channels. The gripper needs at least 4 adjacent 1000 µL channels0.7.0
CO-RE 96 Multi-Probe HeadCore 96 ServiceThe arm reports a CO-RE 96 head (H0), or has_core96 is true0.7.0

Before connector version 0.5.0, the connector registered all services regardless of the installed modules.

The Microlab VANTAGE is a complex instrument, and direct commands can cause hardware collisions. Always use the Liquid Handling SDK for operations that move the instrument.

First successful test

The connector is a SiLA server and accepts direct calls. Connect the SDK to the Platform, then run this code in a notebook or an async script. It reads the firmware version and doesn't move the instrument.

Replace Hamilton Microlab Vantage with the service name shown in GroundControl. Hamilton Microlab Vantage is the default service name of this connector. After a rename, the lookup returns None and the next line fails.

from unitelabs.sdk import AsyncApiClient

client = AsyncApiClient()
vantage = await client.get_service_by_name(name="Hamilton Microlab Vantage")

print(await vantage.gateway_service.get_firmware_version())

Expected output: the firmware version string of the instrument.

Run the same test with the Liquid Handling SDK

Direct connector calls are enough to check a connection. For everything else, use the Liquid Handling SDK. It models the deck, the channels, and the labware, and validates parameters before anything moves. This code reads the firmware version and the instrument configuration through it. As in the previous example, use the service name shown in GroundControl.

from unitelabs.liquid_handling.hamilton import Vantage

vantage = Vantage(name="Hamilton Microlab Vantage")

print(await vantage.get_firmware_version())

configuration = await vantage.get_configuration()
print(configuration)

Expected output: the firmware version string of the instrument, followed by the instrument configuration the SDK reads from the firmware.

Next steps with the Liquid Handling SDK:

Known issues

  • Connector stops at startup. If the connection can't be opened at all, check host, port, and the network connection to the instrument. If the connection opens but the firmware version query fails, for example with a timeout, the connector stops with Device did not respond to firmware version query followed by the reason.
  • A service is missing. At startup, the connector queries the instrument for its arm components and system modules. If a query fails, the connector doesn't stop and doesn't report an error. It writes a debug message to the log, and the services that depend on that query don't appear. A failed arm query removes the PIP, Core Gripper, IPG, XL, and Core 96 Services. A failed system modules query removes the IDL Service. To fix this error, restart the connector and check the network connection to the instrument. If the Core 96 Service is still missing, set has_core96 to true.
  • Firmware timeout of 5 minutes. The Hamilton VANTAGE has a strict limit of 5 minutes (300 seconds) per command, and a command that runs longer is ended by the instrument. The connector can't change this limit, and a longer connector timeout doesn't extend it. Split long running actions, such as complex mixing, into several commands.

Vendor documentation

The Hamilton manuals shipped with the instrument cover operating limits, deck layout, maintenance, and safety.