Connect an Inheco ODTC thermal cycler through GroundControl, check that it answers, and read its temperatures and stored methods from Python.

This page gets you from an unboxed ODTC to a first temperature reading from Python. Inheco's manual covers plate types, method parameters, and safety.

What you need

ItemDetail
InstrumentAn ODTC 96, 96XL, 384, or 384XL, powered on
NetworkEthernet, on the same subnet as the GroundControl machine. If commands time out, see known quirks
Device serial numberThe node name (ODTC_ plus 6 character serial number) from the device label
ConnectorInheco ODTC from the UniteLabs registry
Vendor softwareNone
MachineWindows, Linux, or macOS running GroundControl
AccessA tenant and the SDK installed. See Setup

Install and configure the connector

  1. In GroundControl, add a connector and pick Inheco ODTC (how to add a connector).
  2. Fill in the configuration. The fields that matter:
    FieldValue
    node_nameThe node name from the device label, e.g. ODTC_04E964. Always set it; the default is only an example
    ip_addressLeave empty. Set it to the ODTC's IP address only if the connector can't find the ODTC by name
    simulatedfalse for the real device (true runs the connector's built-in simulator)
  3. Start the connector. It waits until the ODTC reports idle, resetting it first if it's in an error state. GroundControl then shows the modules DeviceService, DoorController, MethodService, and TemperatureProvider.

First successful test

Read the device identity, its state, a set of temperatures, and the methods stored on it. This test does not move or heat up the device.

from unitelabs.sdk import AsyncApiClient

client = AsyncApiClient()
odtc = await client.get_service_by_name(name="Inheco ODTC")

print(await odtc.device_service.get_identification())

status = await odtc.device_service.subscribe_status()
async with status:
    print(await anext(status))

temperatures = await odtc.temperature_provider.subscribe_temperature()
async with temperatures:
    print(await anext(temperatures))

print(await odtc.method_service.get_method_names())

Expected:

  • The identification prints the device name, serial number, and firmware version.
  • The status prints {'State': 'idle', 'Locked': ...}.
  • The temperatures print in °C, keyed Ambient, Heatsink, HeatsinkTEC, Lid, LidMonitor, Mount, and MountMonitor. Mount is the block.
  • The method names print as 2 lists, PreMethods and Methods.

The SDK returns these as plain dicts with PascalCase keys.

Known quirks and errors

You seeIt meansDo this
Startup fails with No ODTC devices were found at allNo ODTC answered discovery: it's off, on another subnet, or the network drops broadcastsCheck power and cabling, and that both are on the same subnet. Then set ip_address
The configuration is rejected on node_nameIt must be ODTC_ plus 6 characters, 0 to 9 and A to FCopy it from the label, in capitals
Commands time out with Timed out waiting for ResponseEventThe ODTC reports finished commands back to the connector, and a firewall on the GroundControl machine blocks itSet event_listener_port to a fixed port, e.g. 8090, and allow inbound TCP on it. On Windows the connector adds this rule itself when it runs with admin rights
returnCode=4The ODTC is busy with another commandWait until the status is idle
returnCode=9The command isn't allowed in the current state, usually after an errorRun device_service.reset()
start fails and the message names StartBlockTemperatureA method's start block temperature has to match the last pre-method's target, or the last method's endRun the matching pre-method first
Method '…' not foundNo pre-method or method on the device has that nameCheck get_method_names(). Names are case-sensitive
start reports progress and finishes, but the ODTC is still running, or it finished earlierstart times progress and its own end from the method's estimated durationWait for the status to return to idle before the next step
A temperature or status subscription stops sending valuesA poll failed, for example when the ODTC finished a command with a warning (returnCode=12), and the connector stops polling after the first errorClose the subscription, then open a new one

More snippets

Open and close the door:

await odtc.door_controller.open()
await odtc.door_controller.close()

Create a method: a method runs from a starting block and lid temperature, and a pre-method brings the ODTC to exactly those temperatures first. Upload both in a single MethodSet in Inheco's XML format. This one holds 20 °C for 90 seconds, then cools to 4 °C and holds for 60 seconds, with the lid at 105 °C.

method_set = """<?xml version="1.0" encoding="utf-8"?>
<MethodSet>
  <DeleteAllMethods>false</DeleteAllMethods>
  <PreMethod methodName="Demo-Pre" creator="labuser" dateTime="2026-10-28T12:00:00">
    <TargetBlockTemperature>20</TargetBlockTemperature>
    <TargetLidTemp>105</TargetLidTemp>
  </PreMethod>
  <Method methodName="Demo-Method" creator="labuser" dateTime="2026-10-28T12:00:00">
    <Variant>960000</Variant>
    <PlateType>0</PlateType>
    <FluidQuantity>2</FluidQuantity>
    <PostHeating>true</PostHeating>
    <StartBlockTemperature>20</StartBlockTemperature>
    <StartLidTemperature>105</StartLidTemperature>
    <Step>
      <Number>1</Number>
      <Slope>4.4</Slope>
      <PlateauTemperature>20</PlateauTemperature>
      <PlateauTime>90</PlateauTime>
      <OverShootSlope1>0.1</OverShootSlope1>
      <OverShootTemperature>0</OverShootTemperature>
      <OverShootTime>0</OverShootTime>
      <OverShootSlope2>0.1</OverShootSlope2>
      <GotoNumber>0</GotoNumber>
      <LoopNumber>0</LoopNumber>
      <PIDNumber>1</PIDNumber>
      <LidTemp>105</LidTemp>
    </Step>
    <Step>
      <Number>2</Number>
      <Slope>0.18</Slope>
      <PlateauTemperature>4</PlateauTemperature>
      <PlateauTime>60</PlateauTime>
      <OverShootSlope1>0.1</OverShootSlope1>
      <OverShootTemperature>0</OverShootTemperature>
      <OverShootTime>0</OverShootTime>
      <OverShootSlope2>0.1</OverShootSlope2>
      <GotoNumber>0</GotoNumber>
      <LoopNumber>0</LoopNumber>
      <PIDNumber>1</PIDNumber>
      <LidTemp>105</LidTemp>
    </Step>
    <PIDSet>
      <PID number="1">
        <PHeating>60</PHeating>
        <PCooling>80</PCooling>
        <IHeating>250</IHeating>
        <ICooling>100</ICooling>
        <DHeating>10</DHeating>
        <DCooling>10</DCooling>
        <PLid>100</PLid>
        <ILid>70</ILid>
      </PID>
    </PIDSet>
  </Method>
</MethodSet>"""

await odtc.method_service.set_parameters_as_xml(params_xml=method_set, delete_all=False)
print(await odtc.method_service.get_method_names())  # Demo-Pre and Demo-Method are listed

Remember:

  • Give the pre-method the same block and lid temperatures as the method's StartBlockTemperature and StartLidTemperature. Otherwise method start fails.
  • Leave optional attributes such as description out of the XML, or give them a value. The ODTC rejects empty ones, such as description="".

delete_all=False keeps the methods already on the ODTC. It does not override a <DeleteAllMethods>true</DeleteAllMethods> in the XML. Repeat steps: to loop, as in a PCR cycle, set GotoNumber on the last step of the loop to the step it jumps back to, and LoopNumber to the number of extra passes.

Run a method and watch the block: run the pre-method with the door closed, load the plate, then run the method. start blocks for the method's estimated duration, so run it as a task and read temperatures alongside it. Then wait for idle, since the device can still be busy when start returns.

import asyncio


async def run_and_watch(name):
    run = asyncio.create_task(odtc.method_service.start(method=name, priority=1))
    temperatures = await odtc.temperature_provider.subscribe_temperature()
    async with temperatures:
        while not run.done():
            reading = await anext(temperatures)
            print(f"{name}: block {reading['Mount']} °C, lid {reading['Lid']} °C")
    await run

    status = await odtc.device_service.subscribe_status()
    async with status:
        while (await anext(status))["State"].lower() != "idle":
            pass


await odtc.door_controller.close()
await run_and_watch("Demo-Pre")

await odtc.door_controller.open()
input("Load the plate, then press Enter")
await odtc.door_controller.close()

await run_and_watch("Demo-Method")

Read the trace of the last run: await odtc.temperature_provider.get_last_data() returns a sample per time point, with target, block, and lid temperatures.

Manage methods: get_parameters_as_xml() returns everything on the ODTC in the same XML format, delete_method(name=...) removes one, and delete_all=True on upload replaces them all.

Connector reference

Browse the connected instrument's modules and actions in GroundControl, or list them in Python with print(odtc.method_service.actions.keys()).

Vendor documentation

Plate types, method parameters (slopes, overshoot, PID sets), and safety are in Inheco's ODTC manual.

Last updated