QInstruments BioShake
This page gets you from an unboxed BioShake to a first shake driven from Python. It does not replace the QInstruments manual: speeds, loads and safety are the vendor's domain.
unitelabs-qinstruments-bioshake 0.3.0). It gains a "tested against" label once the
snippet pipeline lands.What you need
| Item | Detail |
|---|---|
| Instrument | Any BioShake model listed above, powered on, plate adapter mounted |
| Cable | The USB (RS-232) cable QInstruments ships; the device appears as a serial port |
| Connector | QINSTRUMENTS BioShake from the UniteLabs registry — one connector drives every QInstruments model, including the ColdPlate, TiltStation and TurnStation |
| Vendor software | None. Close any QInstruments or third-party tool that holds the serial port |
| Machine | Windows, Linux or macOS running GroundControl |
| Access | A tenant and the SDK installed — see Setup |
Install and configure the connector
- In GroundControl, add a connector and pick QINSTRUMENTS BioShake from the registry; GroundControl downloads the build for your operating system (how to add a connector).
- Fill in the configuration. The fields that matter:
Field Value deviceYour exact model, e.g. BioShake 3000-T elm. The list holds every QInstruments device this connector supports — pick the BioShake entry, not TiltStation or TurnStation, even if they appear firstserial_numberThe serial printed on the device label (max 10 characters); leave empty to skip the identity check portThe serial port the BioShake appears on, e.g. COM3or/dev/ttyUSB0simulatedfalsefor the real device (trueruns the connector's built-in simulator) - Start the connector. In GroundControl it shows the modules
ConfigurationService,ShakeController,TemperatureController(T variants) andElmController/ClampController(elm variants). If it reportsDeviceIdentityError, the port or the serial number is wrong.
First successful test
A ten-second shake at a gentle speed, then a state read-back. Run it in a notebook or an async
script with the SDK authenticated (Connect to Platform).
import asyncio
from unitelabs.sdk import AsyncApiClient
client = AsyncApiClient()
shaker = await client.get_service_by_name(name="QINSTRUMENTS BioShake")
# Identify what is connected and make sure it is awake (Eco Mode rejects every command).
print(await shaker.configuration_service.get_device_info())
await shaker.configuration_service.leave_eco_mode()
# Shake for 10 s at 500 rpm with a 5 s ramp, then confirm the shaker is back home.
await shaker.shake_controller.start_shaking_with_runtime(runtime=10, target_speed=500, acceleration_time=5)
await asyncio.sleep(12)
print(await shaker.shake_controller.get_shaker_state_with_description())
Expected: the device info line names your model and firmware; the plate shakes for ten seconds; the final state reads as stopped in the home position.
Known quirks and errors
| You see | It means | Do this |
|---|---|---|
EcoModeActive | The device is in Eco Mode and rejects every command | await shaker.configuration_service.leave_eco_mode() first; disable Eco Mode on the device if it keeps returning |
DeviceIdentityError | Wrong serial number in the configuration, or no device on the port | Check port; clear serial_number to skip the identity check |
DeviceSelftestException — "activate the boot screen text first" | Self-test results are only readable when the boot screen text is on | await shaker.configuration_service.enable_boot_screen(), then reset_device_with_self_tests() |
ShakerError (code 103) | The shaker was not initialised after power-on | Power-cycle, then await shaker.configuration_service.reset() before shaking |
ElmAlreadyLocked / ElmAlreadyUnlocked | The lock is already in the requested state | Nothing — the call is idempotent in effect; read get_elm_state_as_string() first if you need to branch |
ReleaseBlocked | The clamp cannot release while the shaker moves | stop_shaking() (waits for home position), then clamp_controller.release() |
OverheatError | Over-temperature protection tripped | Let the device cool; on code 33020 power-cycle it |
| GroundControl shows "TiltStation" for a BioShake | The device field was left on the first list entry | Set device to your BioShake model and restart the connector |
Speed limits differ per model: read them with get_device_specific_minimum_shaking_speed() and
get_device_specific_maximum_shaking_speed() before choosing a target.
More snippets
- Temperature (T variants):
set_target_temperature(target_temp=37.0), thenturn_temperature_control_on(); follow it withsubscribe_current_temperature(). - Hold the plate:
elm_controller.lock_elm()before shaking a loose plate,unlock_elm()afterstop_shaking(). - Emergency stop:
shake_controller.emergency_stop()stops at an undefined position — follow withmove_to_home_position()before a gripper approaches.
Connector reference
print(shaker.modules.keys()) and print(shaker.shake_controller.actions.keys()).Vendor documentation
Operating limits, plate adapters, maintenance and safety are covered by QInstruments' BioShake operating manual shipped with the device. This page only covers getting it connected and tested.