Get a QInstruments BioShake (3000, 3000-T, 5000, D30, Q1, Q2 and their elm variants) connected through GroundControl and run a first shake from Python.

This page gets you from an unboxed BioShake to a first shake driven from Python. It does not replace the QInstruments manual: speeds, loads and safety are the vendor's domain.

The code on this page is not yet run by the docs pipeline; it follows the connector's public interface (unitelabs-qinstruments-bioshake 0.3.0). It gains a "tested against" label once the snippet pipeline lands.

What you need

ItemDetail
InstrumentAny BioShake model listed above, powered on, plate adapter mounted
CableThe USB (RS-232) cable QInstruments ships; the device appears as a serial port
ConnectorQINSTRUMENTS BioShake from the UniteLabs registry — one connector drives every QInstruments model, including the ColdPlate, TiltStation and TurnStation
Vendor softwareNone. Close any QInstruments or third-party tool that holds the serial port
MachineWindows, Linux or macOS running GroundControl
AccessA tenant and the SDK installed — see Setup

Install and configure the connector

  1. In GroundControl, add a connector and pick QINSTRUMENTS BioShake from the registry; GroundControl downloads the build for your operating system (how to add a connector).
  2. Fill in the configuration. The fields that matter:
    FieldValue
    deviceYour exact model, e.g. BioShake 3000-T elm. The list holds every QInstruments device this connector supports — pick the BioShake entry, not TiltStation or TurnStation, even if they appear first
    serial_numberThe serial printed on the device label (max 10 characters); leave empty to skip the identity check
    portThe serial port the BioShake appears on, e.g. COM3 or /dev/ttyUSB0
    simulatedfalse for the real device (true runs the connector's built-in simulator)
  3. Start the connector. In GroundControl it shows the modules ConfigurationService, ShakeController, TemperatureController (T variants) and ElmController / ClampController (elm variants). If it reports DeviceIdentityError, the port or the serial number is wrong.

First successful test

A ten-second shake at a gentle speed, then a state read-back. Run it in a notebook or an async script with the SDK authenticated (Connect to Platform).

import asyncio

from unitelabs.sdk import AsyncApiClient

client = AsyncApiClient()
shaker = await client.get_service_by_name(name="QINSTRUMENTS BioShake")

# Identify what is connected and make sure it is awake (Eco Mode rejects every command).
print(await shaker.configuration_service.get_device_info())
await shaker.configuration_service.leave_eco_mode()

# Shake for 10 s at 500 rpm with a 5 s ramp, then confirm the shaker is back home.
await shaker.shake_controller.start_shaking_with_runtime(runtime=10, target_speed=500, acceleration_time=5)
await asyncio.sleep(12)
print(await shaker.shake_controller.get_shaker_state_with_description())

Expected: the device info line names your model and firmware; the plate shakes for ten seconds; the final state reads as stopped in the home position.

Known quirks and errors

You seeIt meansDo this
EcoModeActiveThe device is in Eco Mode and rejects every commandawait shaker.configuration_service.leave_eco_mode() first; disable Eco Mode on the device if it keeps returning
DeviceIdentityErrorWrong serial number in the configuration, or no device on the portCheck port; clear serial_number to skip the identity check
DeviceSelftestException — "activate the boot screen text first"Self-test results are only readable when the boot screen text is onawait shaker.configuration_service.enable_boot_screen(), then reset_device_with_self_tests()
ShakerError (code 103)The shaker was not initialised after power-onPower-cycle, then await shaker.configuration_service.reset() before shaking
ElmAlreadyLocked / ElmAlreadyUnlockedThe lock is already in the requested stateNothing — the call is idempotent in effect; read get_elm_state_as_string() first if you need to branch
ReleaseBlockedThe clamp cannot release while the shaker movesstop_shaking() (waits for home position), then clamp_controller.release()
OverheatErrorOver-temperature protection trippedLet the device cool; on code 33020 power-cycle it
GroundControl shows "TiltStation" for a BioShakeThe device field was left on the first list entrySet device to your BioShake model and restart the connector

Speed limits differ per model: read them with get_device_specific_minimum_shaking_speed() and get_device_specific_maximum_shaking_speed() before choosing a target.

More snippets

  • Temperature (T variants): set_target_temperature(target_temp=37.0), then turn_temperature_control_on(); follow it with subscribe_current_temperature().
  • Hold the plate: elm_controller.lock_elm() before shaking a loose plate, unlock_elm() after stop_shaking().
  • Emergency stop: shake_controller.emergency_stop() stops at an undefined position — follow with move_to_home_position() before a gripper approaches.

Connector reference

The full list of modules and actions with a copy-paste line per method is generated from the connector's release data in a later step. Until then, browse the modules in GroundControl or with print(shaker.modules.keys()) and print(shaker.shake_controller.actions.keys()).

Vendor documentation

Operating limits, plate adapters, maintenance and safety are covered by QInstruments' BioShake operating manual shipped with the device. This page only covers getting it connected and tested.

Last updated